Flame (AAS) Atomic Absorption Spectrophotometer

<< Previous in Atomic Absorption Next in Atomic Absorption >>
List Price: $20,350.00
Clarkson Price: $18,315.00
You Save: $2,035.00 (10 %)
Clarkson Catalog Number: A3F
Manufacturer: Persee Analytics
Supplier Catalog Number: A3F
Manufacturer Direct Shipment
Persee A3F, Persee A3G, Persee A3FG
Clarkson A3F
Flame (AAS) Atomic Absorption Spectrophotometer
Persee Analytics A3F
 

The A3 Atomic Absorption Spectrometer is a high performance automated instrument designed to meet the requirements of the modern laboratory. Due to its versatility and performance it can be used for a wide range of applications. 

Three flame options are available to the user with the Air/Acetylene being the standard configuration. This flame can be used for nearly all standard elements while the N2O/Acetylene and the Air/LPG (natural gas) are available as an option for the more demanding of elements. All three flame configurations offer coded burner for full safety protection. 

The graphite head is fixed into the optical path to maximize performance and eliminate drift. The temperature of the transversely heated graphite tube is accurately controlled by means of a precision feedback system. Pyrolytically coated platform tubes are supplied as standard to improve the performance and eliminate many analytical problems associated with this technique. 

The flame and graphite furnace can be changed over by a simple selection in versatile AA Win Pro software.

Key Features

  • Easy to use, complete software control
  • Integrated safety precautions for user protection
  • Flexible configuration for a wide array of industrial apps
  • Time-saving high speed analysis
 
Specifications
Spectrometer System
Wavelength range 190-900 nm
Light source Hollow cathode lamp (HCL), Deuterium Arc lamp (D2)
Modulation Square wave pulse
Modulation frequency

100 hz self reversal (SR) background,
400 hz Deuterium Arc (D2) background

Monochromator 
Grating 1800 grooves/mm diffraction grating
Blazed wavelength 250 nm
Focus 300 nm
Bandwidth 0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm
Scan mode Automatic
Photometric type Single beam
Wavelength accuracy ±0.15 nm
Wavelength resolution 0.2 nm±0.02 nm
Wavelength reproducibility <0.05 nm
Spectral bandwidth 0.5, 1.0, 2.0, 5.0 nm
Baseline stability 0.005 A/30 min
Stray light ≤0.12 %T (220 nm NaI, 340 nm NaNO2)
Background correction

Deuterium Arc (D2) 1.0 Abs,
Self Reversal (SR) 3.0 Abs

Flame Analysis
Flame types

Air/Acetylene,
Nitrous Oxide/Acetylene,
Air/Propane (LPG)

Sensitivity (Cu) 2 μg/ml Absorption >0.280 Abs
Characteristic concentration

Cu<0.02 μg/ml,
Ba<0.15 μg/ml (N2O/Acetylene)

Detection limit Cu<0.004 μg/ml
Repeatability Cu<0.7 % (Air/Acetylene Flame)
Ba<1.0 % (Nitrous Oxide/Acetylene Flame)
Burner heads Titanium alloy
Nebulizer

High efficiency glass,
Acid proof available as an option

Pre-mix chamber Corrosion resistant
Atomizer selection Manual (A3F)
Automatic changeover (A3FG)
Safety features Burner identification, flame sensor, gas leak sensor, low gas pressure sensor, drain trap sensor, power loss protection
Graphite Furnace Analysis 
Graphite head Transversally heated
Temperature range Ambient-2650 ℃
Heating program Up to 10 steps. Drying, Ashing, Atomization, Cleaning
Feedback Voltage and optical temperature control feedback
Sensitivity (Cu) 50 ng/m Absorption>0.40 Abs
Detection limit  Cd<0.004 ng/ml
Repeatability

Cu<2.0 %,
Cd<2.0 %

Graphite tubes Pyrolytically coated with platform
Sample size Up to 20μl
Safety features

Burner Identification,
Flame Sensor,
Gas leak Sensor,
Low Gas Pressure Sensor,
Drain Trap Sensor,
Power Loss Protection,
Circulation Water (graphite),
Over Temperature Sensor (graphite)

Data Processing
Operating program AA Win Pro software
Analytical methods

Flame AA, Flame AE,
Graphite Furnace,
Hydride Generation

Readout Mode

Continuous, manual, peak height, peak area

calibrations

Multi-standard calibration, standard addition, interpolation

Data storage

Analytical results, instrument and measurement parameters, signal profile, calibration curve

Power Requirements
Main unit

220 VAC 50/60 Hz 200 W

Graphite power supply

380VAC 50/60Hz,
instantaneous power 5 kW current rating 7 kW

Dimensions 
Main unit

44 x 21 x 21 in
165 lb

Graphite power supply

20 x 21 x 21 in
154 lb

 
Catalog Number Model Number Description
A3F * 500F

Flame (AAS) Atomic Absorption Spectrophotometer. A
data coded Air/Acetylene Burner is included

A3G 500G

Graphite Furnace (GFAAS) Atomic Absorption
Spectrophotometer

A3FG 500FG

Combined Flame & Graphite Furnace Atomic
Absorption Spectrophotometer. A Data Coded
Air/Acetylene Burner is included.

 
Additional Information

Product Reviews

(0 Ratings, 0 Reviews)