High Temperature Furnace 1650°C (3002°F) 3.8 Liter Capacity with CAL9500 Controller
List Price: $13,182.00
Clarkson Price: $12,522.90
You Save: $659.10 (5 %)
Clarkson Catalog Number: XF1-CAL9500
Manufacturer: Vecstar
Manufacturer Catalog Number: XF1-CAL9500
|
|
Clarkson XF1-CAL9500
High Temperature Furnace 1650°C (3002°F) 3.8 Liter Capacity with CAL9500 Controller
Vecstar XF1-CAL9500
Specifications:
| Model
|
XF (1700ºc)
|
XF (1800ºc)
|
|
|
XF1
|
XF2
|
XF3
|
XF4
|
XF5
|
XF6
|
| Chamber Size (m/m)
|
H
W
D
|
140
135
200
|
165
160
270
|
200
200
350
|
140
135
200
|
165
160
270
|
200
200
350
|
| External Size (m/m)
|
H
W
D
|
820
570
770
|
870
570
815
|
950
640
890
|
820
570
880
|
870
570
915
|
950
640
950
|
| Volume (Litres)
|
3.8
|
7.1
|
14
|
3.8
|
7.1
|
14
|
| Maximum Temperature
|
1700ºC
|
1800ºC
|
| Maximum Continuous WorkingTemperature
|
1650ºC
|
1750ºC
|
| Temperature Sensor
|
Type 'B'
|
Pt 20% Rh / Pt 40% Rh
|
| Max Power Rating (kw)
|
5.5
|
6.6
|
9.0
|
5.5
|
7.6
|
11.5
|
| Holding Power (kw)
|
2.2
|
2.64
|
3.6
|
2.2
|
3.0
|
4.6
|
| Net Weight (kg)
|
|
|
|
|
|
|
| Electrical Supply
|
1 PH
|
| Power Control
|
Solid State Relay
|
| Overtemperature Protection
|
Independent, Customer Programmable
|
| Insulation Materials
|
Graded Fibre Board
|
| Controller
|
See below for available options
|
|
Controller Specifications:
|
CAL9500 Controller *
The 48x48mm controller offers simple user control.
Setpoint and actual measured temperature are displayed simultaneously.
A 126 segment programmer allocated in up to 31 programs including guaranteed dwell.
West ProVU Programmer
The text based controller is available in multilanguages and offers simple user control.
Setpoint and actual measured temperature are displayed simultaneously. An onscreen trend function shows historic data and dual colour LED backlight changes on alarm condition for improved visibility.
A 255 segment programmer allocated in up to 64 programs including guaranteed dwell.
Optional Extras
Over-temperature protection
West 6700 48x48mm limit controller protects the furnace and its contents at any specified temperature.
Digital Communications
RS232 allowing a single controller to communicate with a computer .
RS485 allowing multiple controllers to communicate by "Daisychaining"
Software
Software packages are available for controller configuration and monitoring.
Cascade Control
Available when close monitoring of the work load is important. An additional thermocouple is placed close to the load and a second controller is fitted. The thermocouple sends signals to the normal temperature controller which in turn adjusts the element temperature.
Chart Recorders
A wide selection of both analogue and graphic recorders are available on request.
Calibration Certification
All controllers and chart recorders can be supplied with three point calibration (customer choice) certificate supplied by the controller manufacturere
Please note all other control manufacturers systems available on request.
|
|
|
|
|