Waterproof pH/EC/TDS/Temperature Meter with pH Electrode Diagnostics
Clarkson Price: This item has been upgraded to HI9813-6N
Clarkson Catalog Number: HI9813-6
Manufacturer: Hanna Instruments
Manufacturer Catalog Number: HI9813-6
|
|
Clarkson HI9813-6
Waterproof pH/EC/TDS/Temperature Meter with pH Electrode Diagnostics
Hanna Instruments HI 9813-6
Description:
pH electrodes are commonly designed to produce and output signal equal to 0 mV when immersed in a solution with pH 7.
But a constant output value at pH 7 occurs as a result of certain adverse events, such as:
• A broken bulb in the pH electrode
• A short circuit in the cable or connector
• Magnetic force leakage in the cable or connector (for example, in case of excessive humidity)
In these situations, the instrument displays a false reading at pH 7 and the user may not be aware that corrective measures are necessary; particularly if the user is anticipating a pH neutral reading.
To avoid this problem, HANNA® developed F.S.T. technology for HI 9813-6.
With this special security system, a system failure will force a fixed value close to pH 4, rather than pH 7, and render the problem clearer to the user.
For added confidence in your measurements, the meter is equipped with two LEDs and a "CHECK" button. While pressing this button, upon immersing the electrode in the Cal-Check solution HI 50021P, one of the LEDs will light up to indicate the status of the meter. The green LED ("OK") indicates the probe is working correctly, while the red LED ("C&C" Clean & Cal) indicates the electrode needs attention.
Specifications:
|
Range
|
pH
|
0.0 to 14.0 pH
|
|
Range
|
EC
|
0.00 to 4.00 mS/cm
|
|
Range
|
TDS
|
0 to 1999 ppm (mg/L)
|
|
Range
|
Temperature
|
0.0 to 60.0°C
|
|
Resolution
|
pH
|
0.1 pH
|
|
Resolution
|
EC
|
0.01 mS/cm
|
|
Resolution
|
TDS
|
1 ppm (mg/L)
|
|
Resolution
|
Temperature
|
0.1°C
|
|
Accuracy
|
pH
|
±0.1 pH
|
|
Accuracy
|
EC
|
±2% F.S.
|
|
Accuracy
|
TDS
|
±2% F.S.
|
|
Accuracy
|
Temperature
|
±0.5°C
|
|
TDS Conversion Factor
|
|
0.56 to 0.72 ppm = 1 µS/cm
|
|
Calibration
|
|
manual at 1 point (all parameters except temperature)
|
|
Temperature Compensation
|
|
automatic, 0 to 50°C (32 to 122°F) with ß = 2%/°C (EC/TDS only)
|
|
Probe
|
|
HI 1285-6 pH/EC/TDS/T,
with Clean & Cal, 1 m (3.3') cable (included)
|
|
Battery Type / Life
|
|
1 x 9V / approx. 150 hours of continuous use
|
|
Environment
|
|
0 to 50°C (32 to 122°F); RH max 100%
|
|
Dimensions
|
|
165 x 75 x 45 mm (6.9 x 2.9 x1.8")
|
|
Weight
|
|
180 g (6.3 oz.)
|
Reagents:
|
Electrodes: |
|
|
Special Applications Specific Electrodes for Specific Analysis
|
|
Solutions: |
|
|
Electrode Storage Solution
1 x 500 mL bottle
|
|
|
General Purpose Cleaning Solution
1 x 500 mL bottle
|
|
|
Combination pH/Conductivity Buffer Kit
1413 µS/cm & 7.01 pH
|
|
|
Combination pH/TDS Buffer Kit
1500 ppm & 7.01 pH
|
|
|
Check Solution,25 x 20 mL Sachets
|
|
Accessories: |
|
|
Shockproof Boot
|
|
|
Shockproof Boot
|
|
|
Protective Case for HI 8424NEW
|
Manual: (Download)
Literature: (Download)
|
List Price: $205.00
Clarkson Price: $184.50
You Save: $20.50 (10 %)
Clarkson HI9813-6N Waterproof, Portable pH/EC/TDS/Temperature Meter with CAL CHECK Hanna Instruments HI9813-6N
| | |
|
|