Clarkson Laboratory and Supply Established 1928
We ship world wide

Browse by Manufacturer
Browse Categories

Solution Graphics
 
Search
Search our store

ETS-D6 Electronic Contact Thermometer IKA 3378100

<< Previous in Thermometers Next in Thermometers >>
List Price: $879.00
Clarkson Price: $791.10
You Save: $87.90 (10 %)
Clarkson Catalog Number: 3378100
Manufacturer: IKA
Supplier Catalog Number: 3378100
Manufacturer Direct Shipment
free shipping

Clarkson 3378100
ETS-D6 Electronic Contact Thermometer
IKA 3378100 / IKA 0003378100 
 
Description:
Electronic contact thermometer with 3 patented operating modes of IKA®

- With optimized PID control
- RESET function
- With integrated pH measuring instrument
- Incl. sensor H 62
- For all magnetic stirrers with contact thermometer bushing according to DIN 12878,
  class 2 (e.g. IKA®, Heidolph, and Corning with adapter AD-C1)
- With 3 operating modes
- Ensures perfect temperature control without overshooting the set temperature,
  even in the case of quick heating
- Large, graphic LCD display with multilingual user guide
- Without pH electrode

3 modes of operation guarantee optimum adjustment to your working method:

Operating mode A
Suitable for work with varying parameters (from -50 °C to 450 °C). Safety temperature adjustable

Operating mode B
Suitable for series operation under uniform conditions

Operating mode C
Suitable for unsupervised operation

All values are taken from the memory. This ensures perfect protection against inadvertent improper adjustment

Specifications:
Temperature measuring range min. -50 °C
Temperature measuring range max. 450 °C
Temperature measurement resolution .01 K
Accuracy of temperature measurement ±0.05 + tolerance PT1000 (DIN IEC 751 Class A) K
Heat control accuracy .1 ±K
Control accuracy with sensor .2 ±K
pH measuring range min. 0 pH
pH measuring range max. 14 pH
Accuracy of pH measurement .1 pH
pH measurement resolution .01 pH
Immersion depth max. 200 mm
Permissible ON time 100 %
Dimensions (W x H x D) 96 x 98 x 45 mm
Weight .2 kg
Permissible ambient temperature 0 - 60 °C
Permissible relative moisture 80 %
Protection class according to DIN EN 60529 IP 54
RS 232 interface no
Analog output no
DC Voltage 8 - 16 V=
Power input 15 mA

Product Reviews